Worked example · historical U.S. records
Same P-number. Different changes.
These two records belong to P160017, the original MiniMed 670G application approved September 28, 2016. They illustrate why the S-number matters. They are not a complete history or a determination of today’s labeling.
P160017 / S003
A hardware design change
- Decision
- February 2, 2017 · APPR
- Type in the record
- Normal 180 Day Track
- Reason in the record
- Change Design/Components/Specifications/Material
The statement describes the “Lockout 2.0 Change Design” affecting Guardian Sensor (3), MMT-7020A, B; Guardian Link (3) Transmitter, MMT-7811; and Tester, MMT-776L.
Reading taskCompare the quoted component and model with those named in the change. This statement alone does not establish a new patient population or that a current unit incorporates the change.
Open FDA S003 record
P160017 / S004
A study protocol decision
- Decision
- December 22, 2016 · APPR
- Type in the record
- Normal 180 Day Track No User Fee
- Reason in the record
- Postapproval Study Protocol
The statement says: “Approval of the protocol for the post-approval study (PAS) protocol.”
Reading taskThis is a protocol decision, not a report of study results. Do not describe it as proof that the study was completed, that its outcomes were favorable or that a new model was approved.
Open FDA S004 record
Read dates, not S-number order. S004’s decision (December 22, 2016) predates S003’s (February 2, 2017). Sorting by S-number would give the wrong decision chronology for this pair.
U.S. FDA openFDA PMA records, snapshot October 26, 2026; consulted November 3, 2026. Unit: two historical supplement records, plus the original application for context. Applicant: Medtronic Minimed, Inc., Northridge, California, United States. This location is record context, not a market statistic. U.S. decisions do not establish authorization in another country.